Indian Council of Agricultural Research
(Agricultural Knowledge Management Information System)


ICAR-Central Institute of Fisheries Technology, CIFT Junction Matsyapuri PO, Cochin-682029, Kerala (www.cift.res.in)
Click Here to See Management Committee
Instrument Details
Instrument/Facility Name: LC-MS/MS, 4000 Q Trap, Sciex (2009)
Make                                   : 2009
Model                                  : AB Sciex4000 QTRAPLC-MS/
Specification   :

Click to ViewTurboIonSpray® probe flow rate range 5 to 3000 μL/min APCI probe flow rate range 200 to 3000 μL/min Performance Dynamic Range 1 cps to 4 x 106cps (pulse counting) Scan Type Q1MS, Q3 MS, Product ion, Precursor ion, Neutral Loss or gain, MRM Scan speed User-settable scan speeds of 250 amu/sec, 1000 amu/sec and 4,000 amu/sec in linear ion trap mode. Maximum of 2,400 amu/sec in RF/DC mode NanoSpray™ source with the MicroIonSpray® head flow rate range 0.05 to 2.0 μL/min Mass Range m/z of 5 to 2800 in Q1/Q3 RF/DC mode, m/z of 50 to 2800 in linear ion trap mode NanoSpray™ source using the continuous NanoSpray™ head flow rate 0.05 to 0.5 μL/min NanoSpray™ source with discrete NanoSpray™ head flow rate 0.02 to 0.05 μl/min Mass Stability At vacuum and electronics equilibrium, 0.1 amu over 9 hrs at m/z 2010 with normal operating temperature Mass Resolution >3000 at m/z 609 at 250 amu/sec Physical Specifications Width 130 cm (53 in) Depth 50 cm (20 in) Weight 136 kg (299 lb) Height 52 cm (21.5 in) Computer Requirements (PC) Operating System Windows 2000 CD-ROM 40x/10x/40x CDRW Floppy 1.44 MB Memory 256 MB CPU 2.53 GHz Hard Disk 40 GB Display 18 inches Viewable Flat Panel Monitor recommended Power Requirements Power Consumption 2500VA max Voltage 2 separate 208V minimum to 242 V maximum Frequency Hz 50/60 Hz Circuit Current 15A grounded single-phase circuit

  Working Principles
Click to ViewLC-MS/MS enables quantitative and qualitative analysis within a single system. Use of a quadrupole mass spectrometer as a linear ion trap significantly enhances ion trap performance by increasing ion capacity, improving injection and trapping efficiencies, and increasing duty cycle.The high sensitivity of this instrument ensures low detection limits. Efficient ionization and superior ion sampling of the interface result in excellent sensitivity. Easy-to-use Turbo V Source allows to quantitate over a wide range of flow rates and to change rapidly between atmospheric-pressure chemical ionization (APCI) and TurboIonSpray Probes. The optional DuoSpray Ion Source accelerates method development and increases both throughput and data quality. An additional option for APCI, the PhotoSpray Source, expands the range of compounds for analysis. The software applications enable automated MS to MS/MS acquisition, allows to extract the maximum amount of information from a single LC-MS run.The superior ionization efficiency, enhanced robustness, and ease-of-use of the innovative Turbo V Source, combined with the LINAC Collision Cell give high performance and high productivity with greater confidence.

  Applications
  User Instructions

  Contact Us
Contact No. 0484-2412300
Email Id zynu@rediffmail.com

  LC-MS/MS, 4000 Q Trap, Sciex (2009) Charges
Procedure Name
Industry
University
National LAB/R&D
Remark
Standarization & Estimation
6000
6000
6000
Per component/sample
Subsequent sample per element
6000
6000
6000
6000
Sample Preparation
2000
2000
2000
per sample

Additional Rs 100/-towards postal charges if result are to be posted
The Demand Draft should be in favour of "The Director, ICAR-CIFT, Cochin OR the payment can be made directly at ICAR-CIFT office"
Letter, DD & Samples send to ""