Instrument/Facility Name: Oxygen Transmission Rate (OTR) Analyser
Make                                   : Lyssy PBI Dansensor
Model                                  : Oxygen Permeability Tester OPT 5000
Specification   :This instrument is under NABL scope

Click to View• Measuring range: 0.1-10 cc/m2/day as standard • Sensor type: Ceramic solid state oxide sensor • Carrier gas: Nitrogen with low oxygen content (eg. N2 5.0/ UHP/ White spot grade) • Test gas: Oxygen 1-100% (100% or 20.9% recommended) • Sensor life: > 5 years • Cascade measuring: Automatic sequential run of multiple test programmes

  Working Principles
Click to ViewSample to be tested in the OPT-5000 may be placed in sample card which fixes the sample correctly in the chamber which is located behind the lid.

  Applications
  User Instructions
Dry measurements: Temperature Range: 5-60 oC Humid measurements: 20-90% RH symmetrically both sides on the following temperature range: 20-40 oC, Limited RH ranges at temperatures 10-20 oC and 40-50 oC The sample must be cut to a size of 10 x 10 cm using the supplied template. Fold the card along the folding line, so it folds easily and place it on the table with the inside facing upwards. • Place the sample with the face up above the hole on the short side of the card. The sample must cover the hole; it must not protrude over the edge of the card or the fold. There must be a bit of empty space around the sample. Fold the long side of the card over the sample and press it down over the short side of the card so it is straight. Use a blunt instrument to press along the edges to make the glue stick.

  Contact Us
Contact No. 0484-2412300
Email Id ashok1960@gmail.com

  Oxygen Transmission Rate (OTR) Analyser Charges
Procedure Name
Industry
University
National LAB/R&D
Remark
OTR (cc/m2/day)
1500
1500
1500
per sample

Additional Rs 100/-towards postal charges if result are to be posted
The Demand Draft should be in favour of "The Director, ICAR-CIFT, Cochin OR the payment can be made directly at ICAR-CIFT office"
Letter, DD & Samples send to "The Director, ICAR-CIFT, Willingdon Island, Cochin, Kerala-682029 "